MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
S
PECIFICATIONS
14
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Specificatio
ns
Filtration, Additional
X-ray beam = 75 kV
EN 60601-1-3 par. 7.5
Optional item
Limitation of Extra Focal Radiation
Set Focus Distance, SID 100 cm (40")
EN 60601-2-54
< 150 mm
Square X-ray Field Selection
100 cm (40") SID - (± 1% SID)
EN 60601-2-54
Min: 00 x 00 cm
Max: 48 x 48 cm
Round X-ray Field Selection
1 m SID - (± 1% SID)
EN 60601-2-54
n.a.
Light Field Indicator
Luminosity at 1 m from the focus, field size set at 35x35 cm.
EN 60601-2-54
> 160 lx
Light Field Indicator - Contrast
Edge Contrast Setting 35x35 cm at 1 m SID
EN 60601-2-54
> 4:1
Light Field Indicator Precision
Light Field/X-ray Field Correspondence
EN 60601-2-54
< 1 % SID>
Accuracy of X-ray field as shown on the front display versus actual
X-ray field dimension.
< 2 mm per side>
X-ray Field Indication Precision
Settings on an Index Scale
EN 60601-2-54
< 2% SID>
SID: (optional)
Precision of Measurement with Retractable Tape
EN 60601-2-54
< 2% SID
Maximum Radiation Leakage
Measured at 100 cm with X-ray Beam = 150 kVp - 4 mA
EN 60601-1-3
< 40 mRh
R
ADIOLOGICAL
S
PECIFICATION