• Turn the right knob clockwise to switch from NO to YES.
• Confirm with LIGHT (C) push button to get into the LONG Manual Calibration.
• Rotate the left knob to adjust the LONG shutter to the desired aperture.
• Confirm with LIGHT (C) push button.
• Press FILTER (D) to go to the next aperture.
T
HE
POTENTIOMETER
VALUE
DIFFERENCE
MUST
BE
AT
LEAST
300
STEPS
OTHERWISE
,
THE
SOFTWARE
GIVES
THE
ERROR
MESSAGE
.
CROSS 300 mm P: xxx
CROSS 240 mm P: xxx
CROSS 130 mm P: xxx
CROSS 000 mm P: xxx
LONG Manual Calibr.
LONG 480 mm P: xxx
LONG 400 mm P: xxx
LONG 300 mm P: xxx
LONG 240 mm P: xxx
LONG 130 mm P: xxx
LONG 000 mm P: xxx