Model R 221 ACS - R 221 ACS DHHS
MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
C
AN
B
US
M
ESSAGES
72
MTR
2
21
AC
S
-
MTR 22
1
ACS DHHS - Can Bus Me
ssages
Tx_Con
fig
0x600
8
0
0x53
filterNr
Ch1
Ch2
Ch3
Ch4
Ch5
Ch6
Configure the name of the spectral filter selected.
filterNr : number of the filter for which to configure the name.
Ch1…Ch6 : first bundle of Characters for setting the name of the
spectral filter.
Note : If the spectral filter name is shorter than 6 characters, use
0x00 as terminator.
Otherwise, complete the name with the second bundle of
characters defined with the configuration command 0x600 sub
0x53.
1.13
Tx_Con
fig
0x600
8
0
0x54
filterNr
Ch7
Ch8
Ch9
Ch10
Ch11
Ch12
Configure the name of the spectral filter selected.
filterNr : number of the filter for which to configure the name.
Ch7…Ch12 : second bundle of Characters for setting the name
of the spectral filter.
Note : the last character of the spectral filter name shall be 0x00.
1.13
Tx_Con
fig
0x600
8
0
0x55
0x01
M50
0x00
0x00
0x00
0x00
0x00
Configure the ASR003 M50 pin.
D2 = 0x00 : Open = automatic, close = manual (default)
D2 = 0x01 : Open = manual; close = automatic
1.13
Tx_Con
fig
0x600
8
0
0x55
0x02
M38
0x00
0x00
0x00
0x00
0x00
Configure the ASR003 M38 pin.
D2 = 0x00 : Open = automatic, close = manual (default)
D2 = 0x01 : Open = manual; close = automatic
1.13
Tx_Con
fig
0x600
8
0
0x55
0x03
M56
0x00
0x00
0x00
0x00
0x00
Configure the ASR003 M56 pin.
D2 = 0x00 : If Open the collimator operates in Manual Mode
(Default)
D2 = 0x01 : If Open the collimator operates in Automatic Mode,
and the LCD displays FTD instead of SID.
1.13
Tx_Con
fig
0x600
8
0
0x56
MSB_OutputFreq
LSB_OutputF
req
0x7F3Out
putEnabli
ng
0x00
0x00
0x00
0x00
Configure the GC-LED 4A PCB Diagnostic message.
MSB_OutputFreq, LSB_OutputFreq = output frequency for the
0x7F3 status message. Eligible range : [50; 1000]ms.
If 0, 0x7F3 status message is not retrieved.
0x7F3OutputEnabling = if 1, 0x7F3 status message emitted
from the collimator.
1.14
Tx_Con
fig
0x600
8
0
0x64
SaveConfig
0x00
0x00
0x00
0x00
0x00
0x00
Save the X-ay Collimator configuration parameters with values
set.
SaveConfig = 0x00 : discard modifications.
SaveConfig = 0x01 : save modifications.
1.00
Tx_Con
fig
0x600
8
0
0xFD
0xAA
0x00
0x00
0x00
0x00
0x00
0x00
Restore the X-Ray Collimator configuration parameters to their
default values.
1.09
Tx_Con
fig
0x600
8
0
0xFE
0x55
0x00
0x00
0x00
0x00
0x00
0x00
Reboot the X-Ray Collimator.
1.00
Tx_Con
fig
0x600
8
0
0xFF
EnStepperMotorModif
0x00
0x00
0x00
0x00
0x00
0x00
Enable or disable the X-Ray Collimator stepper motor
configuration paratemer.
EnStepperMotorModif = 0x00 : Disable the stepper motor
configuration parameter.
EnStepperMotorModif = 0x01 :Enable the stepper motor
configuration parameter.
1.00
Tx_Con
fig
0x601
8
0
Mux
0x00
0x00
0x00
0x00
0x00
0x00
0x00
Read the X-Ray Collimator Configuration Value for the
parameter specified by Mux.
1.00
Type
ID
DLC RTR
D0
D1
D2
D3
D4
D5
D6
D7
Description
Fro
m
FW
Ver
sion