MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
141
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
The 10 output relays each have an NO/NC contact and may consequently be adapted to a
number of different configurations. Their use depends on the software applied.
This board also supplies the go-ahead to program some of the system parameters.
Mechanical Installation of the
External Interface Unit
The external box is to be
installed in such a way as to
ensure connection of all the
electric devices present in the
radiological system.
We suggest that the box be
installed at a safe distance from
electromagnetic sources such
as high voltage transformers.
Use the holes on the sides of the
box to perform the installation.
Should it be necessary to install
the board within a rack of the
general system, remove the electronic board from the box prior to proceeding with the
installation of the box itself.
Make sure that each portion of the rack that is to accommodate the box is adequately earthed.
External Unit Labels
Electrical Connection
• Connect the collimator using the connection cable. The cable end with the AMP 201298-1
connector is to be inserted into the connector at the back of the collimator.
N
OTE
POWER
MUST
BE
SUPPLIED
BY
A
SAFETY
TRANSFORMER
COMPLIANT
WITH
CURRENT
NORMS
AND
PROTECTED
BY
AN
FUSE
4 A.