4.1. Potentiometer Input
Indicate the potentiometer to be calibrated (table or stand)
Press LIGHT (C) to confirm the setting.
4.1.1. Calibration of Vertical SID - Table Down
Indications displayed for the calibration of the table analogue signal require the table to be
lowered:
The second line of the display shows the digitized voltage value generated by the table
potentiometer. The operator will lower the table. When reading is stabilized, press LIGHT (C)
to complete the setting. Should the operator wish to keep the value previously stored and
maintain the last calibration, this setting may be avoided by pressing FILTER (D).
Both instances will lead to the next measurement.
Calibration of Vertical SID - Raise Table 25 cm
The operator is required to place the table at the stated height. When the setting is stabilized,
press LIGHT (C) to confirm the setting.
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER (D).
Calibration of Vertical SID - Focus Table Distance
The operator is required to place the table at the stated height. When the setting is stabilized,
press LIGHT (C) to confirm the setting.
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER (D).
Potentiometer Input
Stand Pot/Table Pot
Move Table Down
xxx pts
Set Table at + 25 cm
xxx pts
Focus Table Distance
xxx pts