MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
164
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
22. Calibration of Left Lateral Bucky Cassette
The system starts with the calibration type.
The Encoder Cross (A) will allow to switch from Quick to Full calibration.
Press LIGHT (C) to confirm the setting.
To cancel the procedure press FILTER (D). This keeps the last calibration values and moves
forward to the next calibration phase. The number of cassette sizes to be processed must
be entered in the event of a Full Calibration.
This step is skipped for the Quick Calibration procedure and 2 cassette values will be
processed.
Full Calibration
xx - cassette number from 0 to 10.
Once the number is selected the cassettes need to be calibrated one by one.
At this point the operator may:
• Set the Cross cassette from 13 cm (“) min to 43 cm (17”) max with the Encoder
Cross (A).
• Set the Long cassette from 13 cm (“) min to 43 cm (17”) max with the Encoder
Cross (A).
When values are stabilized, confirm the setting by pressing LIGHT (C).
Left Bucky Cal.
Quick
Set Number of
Left Side Buckie xxx
Cross 1 S cm 13: XXX
Long 1 S cm 13: XXX
Cross 2 S cm 43: XXX
Long 2 S cm 43: XXX