MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
160
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
9. Left Lateral Receptor
The type of lateral receptor (left) installed with the system must be selected. The Encoder
Cross (A) will allow to switch between BUCKY, CAN, NO.
•
NO:
no receptor is connected.
•
BUCKY:
classical Bucky receptor.
•
Left Film-Plane Distance
This requests the distance from the receptor top to the image receptor.
The display will read the following information
The Encoder Cross (A) will allow to enter values from 0 min to 15 cm max.
Press LIGHT (C) to confirm the setting; the next selection will be displayed.
•
CAN:
formats are transmitted to the collimator via CanBus.
Press LIGHT(C) to confirm the setting; depending on the vertical SID value entered.
10. Setting of SID Horizontal # 1
The display will require the fixed values for the lateral SID to be entered.
The Encoder Cross (A) will allow to enter the measurements from 25 min to 300 cm max
Values will be given in inches if this was the initial setting.
Press LIGHT (C) to confirm the setting and move forward to the to the next Lateral SID value.
11. Setting of SID Horizontal # 2
The Encoder Cross (A) will allow to enter the measurements from 25 min to 300 cm max.
Press LIGHT (C) to confirm the setting.
12. Setting of SID Horizontal # 3
The Encoder Cross (A) will allow to enter the measurements from 25 min to 300 cm max.
Press LIGHT (C) to confirm the setting.
Left Lateral Receptor
NO/BUCKY/CAN
Left FilmPlane Distance
xxx cm
Horiz. SID Meas. #1
xxx cm
Horiz. SID Meas. #2
xxx cm
Horiz. SID Meas. #3
xxx cm