4.1.2. Calibration of Vertical SID - Stand All Up
The display will provide calibration indications for the stand analogue signal, requiring the
stand be raised to a higher position.
The second line of the display shows the digitized voltage value generated by the stand
potentiometer.
The operator is required to raise the stand. When the value is stabilized, press LIGHT (C) to
confirm the setting.
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER (D).
Calibration of Vertical SID - Lower Stand 50 cm
The operator is required to place the table at the assigned height. When the value is
stabilized, press LIGHT (C) to complete the setting.
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER (D).
Calibration of Vertical SID - SID at 100 cm
The operator is required to set the stand at 1 m (100 cm) from the table top.
When the value is stabilized press LIGHT (C) to complete the setting
Should the operator wish to keep the value previously stored and maintain the last
calibration, this setting may be avoided by pressing FILTER (D).
4.2. Fixed Vertical SID Value
The Encoder Cross (A) will allow to increase or decrease the SID value from 50 cm min to
300 cm max.
Press LIGHT to confirm the setting; the next selection will be displayed (go to section 5):
Move Stand Full Up
xxx pts
Lower 50 cm
xxx pts
Set stand to 100cm
xxx pts
Vertical Fixed SID
xxx cm