MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
154
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
3.3. Offset Inclinometer - 90° CW
WARNING
M
AKE
SURE
THAT
THE
COLLIMATOR
IS
ROTATED
90°
TOWARD
LEFT
(CW)
WHILE
ACQUIRING
DATA
,
OTHERWISE
THE
CALIBRATION
WILL
NOT
BE
VALID
.
xx: value read by inclinometer
yy: value read by inclinometer
Press LIGHT (C) to confirm the setting.
3.4. Degrees Tolerance
The encoder Cross (A) will allow to enter values from 1 to 15 degrees.
Press LIGHT (C) to confirm the setting.
3.5.
Collimator Angle Visualization
The Encoder Cross (A) activates or deactivates the display of the collimator tilt angle in
degrees.
Press LIGHT (C) to confirm the setting.
4. Vertical SID Measure
The Encoder Cross (A) will allow to switch between SID: SINGLE, DIFFERENTIAL, FIXED
and CAN.
•
SINGLE:
only one potentiometer is calibrated for SID calculation:
• Table (go to the section
4.1
and
4.1.1
)
• Stand (go to the section
4.1
and
4.1.2
)
•
FIXED:
the fixed value is set during calibration (go to the section
4.2
).
•
CAN:
SID value is transmitted to the collimator via CANBus (go to
Right Lateral Meas-
ure section 5
)
•
DIFFERENTIAL:
both potentiometers are calibrated, one for the table and one for the
stand (go to the section
4.1.1
and
4.1.2
).
Inclinometer Offset 90 L
X: xx Y: xx
Degrees Tolerance
XX degrees
Angle Visualization
Yes
Vertical SID
Meas.