MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
158
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
•
M
IN
L
EFT
P
OT
. SID
Minimum distance from a surface of image receptor to the film is to be set up.
nnn
- (value read by potentiometer).
•
M
AX
L
EFT
P
OT
. SID
Maximum distance from a surface of image receptor to the film is to be set up.
nnn
- (value read by potentiometer).
7. Vertical Receptor
The Encoder Cross (A) will allow to switch between NO, BUCKY, CAN, ATS or FIXED
FORMATS.
•
NO:
no receptor is connected.
•
BUCKY:
The receptor is a classical Bucky.
•
Film Table Distance
This requests the distance from the table top to the image receptor.
The display will read the following information:
The Encoder Cross (A) will allow to enter values from 0 min. to 15 cm max.
Press LIGHT (C) to confirm the setting.
•
CAN:
formats are CanBus related to the collimator.
•
ATS:
formats are CanBus related to the collimator (ATS detector mode)
Min.Left Pot. SID: nnn
... cm
Max. Left Pot.SID: nnn
... cm
Vertical Receptor
NO/BUCKY/CAN/ATS/FIXED FORMATS
Film Table Distance
xxx cm