Light Field Adjustment
Transversal Calibration (CROSS)*
If the light-field needs calibration, the mirror
needs to be adjusted as follows:
• Remove the part of the cover necessary
to access the screws, see Chapter -
C
OVER
R
EMOVAL
.
• Loosen the mirror fixing screw
A
(not
remove) and rotate the cam
B
to adjust
the position of the mirror, see
Fig.
Transversal Calibration (CROSS).
• Once you have regulated the mirror
tighten the screw
A
and remount the
cover, see Chapter -
C
OVER
R
EMOVAL
.
N
OTE
P
LEASE
CONSULT
YOUR
PERSONALIZATION
PAGE
TO
PERFORM
CORRECT
TRANSVERSAL
CALIBRATION
IF
THE
MECHANICS
DIFFER
FROM
THE
STANDARD
COLLIMATOR
.
Fig. Transversal Calibration