MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
149
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
Check the Following Whenever the Potentiometer Operates Irregularly
:
• The analogue signal (V DC) from the table Bucky potentiometer sliders must increase with
the increase of cassette format. Use a voltmeter to check this condition between contacts
M99 (+) and M100 (-) for the Long function and, between M95 (+) and M96 (-) for the Cross
function.
• The analogue signal (V DC) from the right lateral Bucky potentiometer sliders must
increase with the increase of cassette format. Use a voltmeter to check this condition
between contacts M115 (+) and M116 (-) for the Long function and between M111 (+) and,
M112 (-) for the Cross function.
• The analogue signal (V DC) from the left lateral Bucky potentiometer sliders must increase
with the increase of cassette format. Use a voltmeter to check this condition between
contacts M107 (+) and M108 (-) for the Long function and between M103 (+) and M104 (-)
for the Cross function.
IN-AN
M89
Cable shielding of potentiometer SID/ table
IN-AN
M90
Potentiometer SID/Table terminal (high-level 5 V DC)
IN-AN
M91
Potentiometer SID/Table slider
IN-AN
M92
Potentiometer SID/Table terminal (low-level 0 V DC)
IN-AN
M93
Cable shielding of potentiometer SID/Table
IN-AN
M94
Cross potentiometer/table Bucky terminal (high-level 5 V DC)
IN-AN
M95
Cross potentiometer/table Bucky slider
IN-AN
M96
Cross potentiometer/table Bucky terminal (low-level 0 V DC)
IN-AN
M97
Cable Shielding of Cross potentiometer/table Bucky
IN-AN
M98
Long pot./table Bucky terminal (high-level 5 V DC)
IN-AN
M99
Long potentiometer/table Bucky slider
IN-AN
M100
Long potentiometer/table Bucky terminal (low-level 0 V DC)
IN-AN
M101
Cable Shielding of Long potentiometer/table
IN-AN
M102
Cross potentiometer/left lateral Bucky terminal (high-level 5 V DC)
IN-AN
M103
Cross potentiometer/ left lateral Bucky slider
IN-AN
M104
Cross potentiometer/ left lateral Bucky terminal (low-level 0 V DC)
IN-AN
M105
Cable Shielding of Cross potentiometer/ left lateral Bucky
IN-AN
M106
Long potentiometer/left lateral Bucky terminal (high-level 5 V DC)
IN-AN
M107
Long potentiometer/left lateral Bucky slider
IN-AN
M108
Long potentiometer/left lateral Bucky terminal (low-level 0 V DC)
IN-AN
M109
Cable Shielding of Long potentiometer/ left lateral Bucky
IN-AN
M110
Cross pot./right lateral Bucky terminal (high-level 5 V DC)
IN-AN
M111
Cross potentiometer/right lateral Bucky slider
IN-AN
M112
Cross potentiometer/ right lateral Bucky terminal (low-level 0 V DC)
IN-AN
M113
Cable Shielding of Cross potentiometer/right lateral Bucky
IN-AN
M114
Long pot./right lateral Bucky terminal (high-level 5V DC)
IN-AN
M115
Long potentiometer/ right lateral Bucky slider
IN-AN
M116
Long potentiometer/right lateral Bucky terminal (low-level 0VDC)
IN-AN
M117
Cable Shielding of Long potentiometer/right lateral Bucky
Signal
Contact
Description