MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
226
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
3. With the tabs fully retracted, tabs in position
1 and 3 which overlap the outer ring must be
tightened completely up to the end stroke
with 0.45Nm of torque (at least 7 turns), see
Fig. Mounting Bracket.
4. With the tabs fully retracted, tabs in position
2 and 4 which overlap the outer ring must be
tightened exactly 5 ½ turns, see
Fig. Mount-
ing Bracket.
5. If you are not able to respect the turns some-
thing is incorrect, repeat above instructions,
if issues persist, please contact Ralco.
6. Ensure the microswitch is completely
engaged by tightening the set screw, see
Fig. Microswitch
.
7. The collimator tabs glide on the flange outer
ring and the collimator rotates towards the X-
ray tube axis. If, however:
• The release force from the position 0°
and the collimator rotation is too low,
the two tabs of the tab 2 and 4 need to
be tightened ½ turn.
• The release force from the position 0° and the collimator rotation is too high, the two
tabs of the tab 2 and 4 need to be loosened ½ turn.
Mounting the Collimator to the Metal Flange (RO 611)
N
OTE
T
HE
MOUNTING
FLANGE
PROVIDED
(
IF
PURCHASED
)
WITH
THE
COLLIMATOR
IS
SUBJECTED
TO
TESTING
PURSUANT
TO
ALL
APPLICABLE
STANDARDS
.
M
OUNTING
B
RACKET
T
ABS
CONFORM
TO
EN60601.
W
ARNINGS
P
URSUANT
TO
APPLICABLE
STANDARDS
, R
ALCO
HAS
TESTED
THE
COLLIMATOR
AND
FLANGE
APPLYING
STATIC
LOADS
. R
ALCO
IS
NOT
IN
A
POSITION
TO
KNOW
THE
DYNAMIC
FORCES
OF
ALL
END
-
USER
SYSTEMS
. I
T
IS
THE
RESPONSIBILITY
OF
THE
END
-
USER
TO
ENSURE
DYNAMIC
FORCES
OF
THE
SYSTEM
DO
NOT
CREATE
A
DANGEROUS
CONDITION
.
Fig. Mounting Bracket
*illustrative purpose only