MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
159
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
•
FIXED FORMATS
:
5 fixed formats are selected
.
•
D
IGITAL
F
ORMAT
# n
The display will read indications to enter the fixed measures of lateral SID. It is
possible to connect, to the external board ASR003, the inputs (max 5) that allow to
enter the fixed formats in the automatic mode.
AA - Cross format. The Encoder Cross (A) will allow to increase or decrease the
SID value.
BB - Long format. The Encoder Long (B) will allow to increase or decrease the SID
value.
Values will be given in inches if this was the initial setting.
Press LIGHT (C) to confirm the setting; the next selection will be displayed.
8. Right Lateral Receptor
The type of lateral (right) receptor installed with the system must be selected.
The Encoder Cross (A) will allow to switch between BUCKY, CAN, or NO.
•
NO:
no receptor is connected.
•
BUCKY:
classical Bucky receptor.
•
Right Film-Plane Distance
This requests distance from the receptor top to the image receptor.
The display will read the following information
The Encoder Cross (A) will allow to enter values from 0 min to 15 cm max.
Press LIGHT (C) to confirm the setting; the next selection will be displayed.
•
CAN:
formats are transmitted to the collimator via CanBus.
Press LIGHT (C) to confirm the setting; the next selection will be displayed:
Formato Digitale # n
AA x BB cm
Right Lateral Receptor
NO/BUCKY/CAN
Right FilmPlane Distance
xxx cm