CONFIGURATION
Calibration Start-Up
The collimator can be manually configured with the
use of the display and the control panel.
To start the collimator configuration:
• Switch the collimator ON. The display will show
a series of self-tests.
• Once the self-test is finished press both the
LIGHT (C) and FILTER (D) push buttons and
keep them pressed for approximately
5 seconds.
• Release them when the 3 LEDs light up on the
front panel. See
Fig. 1.
• The display will read:
• Turn the left knob counterclockwise to skip from one menu to the next. There are the four
menus available on the collimator:
RO 525 Customized Knob design
The customer can specify the knob design.
RO 526 Semiautomatic mode (front panel with push buttons included)
This option allows the customer to control the collimator in the semi-automatic mode through
customized software and the front panel.
Wait command . . . .
Collimator Config
Confirm? --> NO
--> In this menu the operator sets all collimator functions
Analog Calibration
Confirm? --> NO
--> In this menu the operator calibrates all potentiometers
present on the X-ray system (Bucky, Stand, Table, Lateral pots)
Shutter Calibration
Confirm? --> NO
--> In this menu the operator calibrates the apertures of LONG
and CROSS shutters
Serial Number
Confirm? --> NO
--> In this menu the operator sets the collimator serial number
A
D
C
B
A - left knob (DSC) changes menus and set up
the Cross cassette size.
B - right knob (DSC) sets up Long cassette size.
C - light push button to confirm
D - filter push button to exit without saving