MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
162
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
18. Calibration of Vertical SID Limits - maximum
The display will show indications of the maximum Vertical SID value accepted by the
collimator.
The Encoder Cross (A) will allow to enter the measurements from 50 cm min to 300 cm max.
Press LIGHT (C) to confirm the setting; maximum Vertical SID data will be displayed.
19. SID Visualisation
The Encoder Cross (A) activates or deactivates the display of the SID value on the main
window.
Press LIGHT (C) to confirm the setting.
20. Key Status Visualization
The Encoder Cross (A) activates or deactivates the display of the key status on the main
window (open or closed padlock).
Press LIGHT(C) to confirm the setting.
Calibrate the vertical Bucky cassettes.
The system will, otherwise, analyze the settings related to the two lateral receptors: if at least
one of the two has been programmed as analogue Bucky, the lateral left and/or right
cassettes will be calibrated.
If no Bucky was selected, the system will display the end of calibration. Cassette Calibration
Information:
The first thing the system requires is to select a quick calibration procedure or a full
calibration procedure for all types of cassettes (vertical, lateral left and lateral right).
The differences between the two calibrations are:
•
Quick Cassette Calibration
: the system requires to enter two cassette sizes only (one
small and one large).
•
Full Cassette Calibration
: the system requires to enter all possible cassette values
(max. 10 Buckys).
IE: If 3 cassette sizes are stored as 13x18, 24x30 and 30x40 cassettes, the system will
also be able to identify cassette sizes such as 13x30, 13x40, 24x18, 24x40, 30x18,
30x30.
Vertical SID Max.
xxx cm
SID Visualization
Yes
Key Status Visual.
Yes