![background image](http://html2.mh-extra.com/html/racal-instruments/3153/3153_user-manual_3135464119.webp)
User Manual 3153
Maintenance and Performance Checks 5-6
Amplitude Accuracy
1. Configure model 3153 to output sine waveforms and a frequency of
1kHz.
*RST;:FREQ 1e3; OUTP ON
2. Perform Amplitude Accuracy tests using Table 5-2.
Table 5-2, Amplitude Accuracy.
DMM Reading
3153 Amplitude Setting Command
Error Limits
CH 1
CH 2
CH 3
Pass
Fail
VOLT 10
3.534V,
±
53.0mV
VOLT 5
1.767V,
±
26.5mV
VOLT 1
353.4mV,
±
5.30mV
VOLT .1
35.34mV,
±
2.35mV
VOLT .01
3.534mV,
±
0.235mV
2 – Low
3 - Low
OUTP:TYPE LOW
2Vp-p
±
100mV
X
Offset
Characteristics
1. Configure model 3153 to output sine waveforms, amplitude of 10.00
mV, and frequency of 1MHz.
2. Perform Offset Accuracy tests using Table 5-3.
3. Precede command settings with INST:SEL 1|2|3 to select the
appropriate channel.
Table 5-3, Offset Accuracy.
DMM Reading
3153 Offset Setting Command
Error Limits
CH 1
CH 2
CH 3
Pass
Fail
VOLT:OFFS 4
±
60mV
VOLT:OFFS 0.4
±
6mV
VOLT:OFFS -0.4
±
6mV
VOLT:OFFS -4
±
60mV
Figure 5-2, Amplitude, Offset, and Square Wave Setup Diagram
REF IN
DMM
+
-
Part 1
Part 2
Part 3
1
2
3
3153
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com