User Manual 3153
SCPI Command Reference 4-12
Standard
Waveform
Command
Summary
The Standard Waveform Commands control the various parameters of
the standard output functions. Optional modes are omitted from these
commands. Factory defaults after *RST are shown in bold typeface.
Parameter low and high limits are given where applicable. The Standard
Waveforms parameters could be used for the APPLy command.
Command and Parameters
Low Limit
High Limit
Default
[:SOURce]
:SINusoid:PHASe <value>
0
360
0
:SINusoid:PHASe?
:TRIangle:PHASe <value>
0
360
0
:TRIangle:PHASe?
:SQUare:DCYCle <value>
1
99
50
:SQUare:DCYCle?
:PULSe:DELay <value>
0
99.9
10.0
:PULSe:DELay?
:PULSe:WIDTh <value>
0
99.9
10.0
:PULSe:WIDTh?
:PULSe:TRANsition <value>
0
99.9
10.0
:PULSe:TRANsition?
:PULSe:TRANsition:TRAiling <value>
0
99.9
10.0
:PULSe:TRANsition:TRAiling?
:RAMP:DELay <value>
0
99.9
0.0
:RAMP:DELay?
:RAMP:TRANsition <value>
0
99.9
60.0
:RAMP:TRANsition?
:RAMP:TRANsition:TRAiling <value>
0
99.9
30.0
:RAMP:TRANsition:TRAiling?
:SINC:NCYCle <value>
4
100
10
:SINC:NCYCle?
:GAUSsian:EXPonent <value>
10
100
10
:GAUSsian:EXPonent?
:EXPonential:EXPonent <value>
-100
100
-20
:EXPonential:EXPonent?
:DC <%_amplitude>
-100
100
100
:DC?