User Manual 3153
Maintenance and Performance Checks 5-8
Table 5-5, Sinewave Characteristics.
Distortion Analyzer Reading
3153 Setting
Reading Limits
CH 1
CH 2
CH 3
Pass
Fail
FREQ 10
< 0.3%
FREQ 100
< 0.3%
FREQ 1E3
< 0.3%
FREQ 1E4
< 0.3%
2 Low
3 Low
FREQ 1E3
< 0.3%
X
FREQ 1E4
< 0.3%
X
FREQ 1E5
< 0.3%
X
Figure 5-3, Sine Characteristics Test Setup Diagram
3. Perform Spectral Purity test using Table 5-5a.
Table 5-5a, Spectral Purity Characteristics.
Spectrum Analyzer Reading
3153 Setting Command
Reading Limits
CH 1
CH 2
CH 3
Pass
Fail
FREQ 1E6
>55dBc
FREQ 10E6|FREQ 5E6
>45dBc
2 Low
3 Low
FREQ 1E6
>55dBc
X
FREQ 10E6|FREQ 5E6
>45dBc
X
REF IN
6900A