User Manual 3153
Using The Instrument 3-28
For the negative EXPonential:
For the GAUSsian:
For the SINC:
where R = N/cyc (number of points per cycle)
cyc = Number of cycles of SINC
Standard waveforms are available at the output connector after the
FUNC:MODE FIX command has been executed.
When using FUNC:SHAP PULS mode, parameters for the PULSe
function shape can be modified, immediately changing the output
waveform. Note that changes made to pulse parameters are
independent of other parameters.
The command:
PULSe:DELay <value>
sets the pulse delay in percent of the pulse period. For example, if the
pulse period is 100 ms, 10% will delay the first transition of the pulse by
10 ms. Delay is measured from trigger to the first transition point.
The query:
PULSe:DELay?
queries the pulse delay setting and returns a value in percent.
The command:
PULSe:TRANsition <value>
sets the pulse rise time in percent of the pulse period. For example, if
the pulse period is 100ms, 5% rise time equals 5ms. Pulse rise time is