User Manual 3153
SCPI Command Reference 4-24
Any of the following commands:
INIT:CONT ON & SEQ:ADV:SING
INIT:CONT OFF & SEQ:ADV:STEP
INIT:CONT OFF & SEQ:ADV:MIX
INIT:CONT OFF;:TRIG:GATE ON & SEQ:ADV STEP
INIT:CONT OFF;:TRIG:GATE ON & SEQ:ADV SING
INIT:CONT OFF;:TRIG:GATE ON & SEQ:ADV MIX
INIT:CONT OFF;:TRIG:BURS ON
Triggered operating mode
Any of the following commands:
OUTP:ECLT0 ON
OUTP:TTLT0 ON
OUTP:TTLT1 ON
OUTP:TTLT2 ON
OUTP:TTLT3 ON
Channel 1, 2 or 3 are separate
An attempt to couple channel 2 or 3 and one of channels 1, 2 or 3 that was
in FIXed function mode.
An attempt to couple channels 2 or 3 to channel 1 while channel 1’s sample
clock frequency is >50 MSa/s and the Channel 2 or 3 clock divider is set to
1.
Channel 2 or 3 are slaves
An attempt to select FIXed function mode on any channel.
An attempt to change sample clock divider to 1 while the channel 1 sample
clock is set to a sample rate >50 MSa/s.
-222,”Data out of range”. Parameter data, which followed a specific header,
could not be used because its value is outside the valid range defined by the
generator.
-224,”Illegal parameter value”. A discrete parameter was received which was not
a valid choice for the command. An invalid parameter choice may have been
used.
-300,”Device-specific-error”. This is the generic device-dependent error for the
instrument when it cannot detect more specific errors. A device- specific error
as defined in IEEE-488.2 has occurred.
-311,”Memory error”. Indicates that an error was detected in the instrument’s
memory.
-350,”Queue Overflow”. The error queue is full because more than 30 errors
have occurred. No additional errors are stored until the errors from the queue
are removed. The error queue is cleared when power has been shut off, or
after a *CLS command has been executed.