User Manual 3153
Specifications
A-6
Triggered
Each input cycle generates a single output cycle.
Internal Trigger
An internal timer repeatedly generates a single output cycle.
Gated
External signal enables generator. First output cycle is
synchronous with the active slope of the triggering signal.
Last cycle of output waveform is always completed.
External Burst
Preset number of up to 1M cycles stimulated by a backplane
TTL trigger or external trigger. This mode is available in
Fixed and User modes only.
Internal Burst
An internal timer repeatedly generates a burst of up to 1Meg
counted output cycles. This mode is available in Fixed and
User modes only. One timer is available for all 3 channels.
Breakpoint
Description
Waveform starts after trigger and stops at breakpoint
position. Each channel can be programmed with its unique
breakpoint
Event to cause stop at breakpoint
SCPI command or valid stop signal at the STOP input.
Breakpoint Error
±2 points
Trigger Start Phase
Description
Waveform starts from point n and completes at point n-1.
Programmable Units
Waveform points, degrees
Range
1 to waveform length; 0
°
to 360
°
Resolution
4 points
Trigger Advance Sources
Internal (On one channel only)
Internal programmable rate generator
Period
From 20
µ
s to 1000s
Accuracy
±(1% + 1
µ
s)
External
Input
1) Front panel SMB
2) VXlbus (TTLTRG 0-3)
3) VXlbus ECLTRG0
Frequency
Fixed and User Mode
DC to 5MHz
Sequence Mode
DC to 500kHz
Software
IEEE 488.2 -TRG, VXI Word Serial trigger command
System Delay
(Trigger to Waveform Output)
Separate
150ns + 1 Sample Clock
Common
150ns + 2 Slowest Sample Clock Periods
Inter-Channel Skew
(Common Trigger Mode)