User Manual 3153
Maintenance and Performance Checks 5-7
Square Wave
Characteristics
1. Configure model 3153 to output a square waveform, amplitude of 8 V,
and frequency of 1 MHz. Run *RST and then:
VOLT 8:;FREQ 1e6;:OUTP ON
2. Use INST:SEL 2 or INST:SEL 3 before the above command string to
set channels 2 and 3.
3. Perform Squarewave Characteristics tests using Table 5-4.
Table 5-4, Squarewave Accuracy.
DMM Reading
3153 Setting
Error Limits
CH 1
CH 2
High
CH 3
High
Pa
ss
Fail
Rise/Fall Time
≤
15ns
Ringing
< 7% + 10mV
Overshoot
< 7% + 10mV
2 Low
3 Low
Rise/Fall Time
≤
5ns
X
Ringing
< 5% + 10mV
X
Overshoot
< 5% + 10mV
X
Sine Characteristics
1. Configure model 3153 to output an arbitrary sine waveform, 4096
points at amplitude of 8 V. Run *RST and then:
VOLT 8;:FUNC:MODE USER;:OUTP ON
Use WaveCAD, Set Axis 4096 points, Wave Sine, Download.
2. Use INST:SEL 1|2|3 before the above command string and download
to select the proper channnel.
3. Perform Sinewave Characteristics tests using Table 5-5.
4. Precede the commands below with INST:SEL 1|2|3 as needed.