The command:
APPLy:SINusoid
{<frequency>,<amplitude>,<offset>,<phase>,<power>}
programs the generator to output a sine waveform with frequency,
amplitude, offset, start phase and power parameters. Parameters are
not optional if the above APPLy command is used. Include all other
parameters in the command. The default settings for these parameters
are: 1MHz, 5Vpk-pk, 0V, 0 and 1.
The command:
APPLy:TRIangle
{<frequency>,<amplitude>,<offset>,<phase>,<power>}
programs the generator to output a triangle waveform with frequency,
amplitude, offset, start phase, and power parameters. The default
settings for these parameters are: 1MHz, 5Vpk-pk, 0V, 0 and 1.
The command:
APPLy:SQUare {<frequency>,<amplitude>,<offset>,<duty_cycle>}
programs the generator to output a square waveform with frequency,
amplitude, offset and duty cycle parameters. The default settings for
these parameters are: 1MHz, 5Vpk-pk, 0V, and 50%.
The command:
APPLy:PULSe{<frequency>,<amplitude>,<offset>,<delay>,
<high_time>,<rise_time>,<fall_time>}
programs the generator to output a pulse waveform with frequency,
amplitude, offset, and delay, rise time, high time and fall time
parameters. The default settings for these parameters are: 1MHz,
5Vpk-pk, 0V, 0%, 10%, 10% and 10%.
The command:
APPLy:RAMP {<frequency>,<amplitude>,<offset>,<delay>,
<rise_time>,<fall_time>}
programs the generator to output a ramp waveform with frequency,
amplitude, offset, delay, rise time, and fall time parameters. The default
settings for these parameters are: 1MHz, 5Vpk-pk, 0V, 0%, 10% and
10%.
The command:
APPLy:SINC
{<frequency>,<amplitude>,<offset>,<number_cycles>}
programs the generator to output a sine(x)/x waveform with frequency,
amplitude, offset, and number of cycles parameters. The default
settings for these parameters are: 1MHz, 5Vpk-pk, 0V and 10.