Clarus 600/680 User’s Guide
3-63
About Sample Vials
Use 2-mL vials with the autosampler (P/N N9301385, 144 per box). These are crimp-top
vials. The vials and crimp caps with septa (P/N 00092205, 1000 per box) are available
through PerkinElmer’s catalog service
(in the U.S., call: 1-800-762-4000) or online at
www. perkinelmer.com
.
The minimum volume of sample required in the 2-mL vial is 0.35 mL or 350
µ
L. You
should fill the vial when running multiple injections per vial.
A mini-vial insert is available for laboratories with small sample volume requirements.
This insert fits into the standard 2-mL vial and requires a minimum volume of 5
µ
L of
sample. These inserts are available through
PerkinElmer’s Catalog service
or visit us on-
line at
www. perkinelmer.com
.
NOTE
: When using mini-vials, it is not recommended to use pumps to acquire samples. This
assures that the sample will not be diluted by residual solvent in the syringe.
Only PerkinElmer vials should be used with the Clarus GC. Adjust the crimper so that it
does not deform the vial cap and septum, but crimps the vial cap tight enough so that you
cannot turn the cap. An improper adjusted crimper can cause the autosampler to
malfunction.
Using a TCD with the Autosampler
When using a TCD at high sensitivity with a GSV or the autosampler, a minor baseline
upset may occur at the time of injection.
If the signal goes below zero, incorporate a timed event Autozero to bring the signal back
to zero before the peaks of interest elute.
Observe the initial chromatogram to determine the proper time for the timed event.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......