Clarus 600/680 User’s Guide
4-89
Optimizing a Detector Signal to an Integrator for
Maximum Sample Dynamic Range
The Clarus GC contains improved electronics which significantly reduce amplifier
noise. This allows you to amplify the detector signal without getting a large
increase in the noise level. Therefore, it provides excellent minimal detection (signal
to noise) for all Clarus GC detectors. Amplification is set on the detector page in
the
Atten
field, the attenuation button, and entering an attenuation value from x1
(most signal) to x64 (least signal). The maximum signal produced by the Clarus GC
amplifier is 1 V. To achieve the best sensitivity and widest dynamic range, the
sample with the highest concentration must not exceed the 1-V signal.
When using a flame ionization detector (FID), if you operate at range x1 and
attenuation x64, and the signal saturates (exceeds 1 V), then use range x20 and
determine the appropriate attenuation value (from x64 to x1) that produces a
reasonable size peak. When using a nitrogen phosphorus detector (NPD), the same
will apply. When using an electron capture detector (ECD), the only range setting is
x1. To ensure that the signal does not exceed 1 V, you should run your most
concentrated sample at attenuation x64, then change the attenuation to produce a
reasonable size peak. When using a thermal conductivity detector (TCD) the range
or current setting also affects the output signal. Therefore, start at attenuation x64.
As long as the signal does not exceed 1 V, you can amplify the signal by changing
the attenuation.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......