Clarus 600/680 User’s Guide
4-5
FID Overview
Flame Ionization Detectors (FID) are mass sensitive and used for the destructive
analysis of organic compounds. As the sample enters the detector, it is mixed with
hydrogen and then burned in air, generating ions. The ions are collected and
measured by the detector, with their concentration proportional to the amount of
compound present. Ion collection is enhanced by a polarized electric field that is
created by applying a negative voltage (
−
200 V) to the jet tip. The detector
sensitivity is affected primarily by the hydrogen flow. The detector linearity is
affected by the air flow.
The FID has
flameout
detection and an
auto-ignite
feature. Flameout detection
occurs 0.5 minutes into the PRE-RUN time and it measures the user-entered
baseline threshold level in mV. If the threshold level is exceeded, the flame is
considered lit, otherwise the Clarus GC detects that the flame is out and auto-ignite
automatically lights the flame before the start of a run. You can also manually light
the flame at any time by pressing the Ignite button
.
A cross section view of a FID is shown below.
Detector
Fitting
Ignitor
Polarizing Pin
Collector
Flame Jet
Extension
Flame Jet
Air In
Hydrogen In
Amplifier
Connector
Figure 4-1 Flame Ionization Detector (FID) cutaway view.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......