Clarus 600/680 GC User’s Guide
6-65
down field is inactive. (GC Tune is only available if it is downloaded. A
successful tune must be performed in order for GC tune to be enabled.
Only those channels for which a valid tune exists will be available on the
drop down list. The selection will be made from the external GC Tune
software, but the user may override this.)
The actual temperature under the single point radio button is the time-
averaged actual oven temperature.
6.
Allow the oven temperature to equilibrate for at least 30 minutes by
setting the time in the
Equilibrate
field. Press
OK
. The displayed
temperature will approach 100.00
°
C (or the oven temperature you
entered in step 4), and the thermometer readout will approach a value
close to 100.00
°
C.
Enter the Required Offset Value
1.
After the temperature has stabilized, subtract the thermometer reading
from the actual oven temperature shown on the previous screen.
2.
Using the plus and minus keys enter a value into the above screen, then
press
OK
. For example, if the thermometer reads 101.55
°
C, and the oven
temperature reads 100.10
°
C, type a value of -1.45
°
C.
Remove the Thermometer Probe
After temperature calibration is satisfactory:
1.
Open the oven door and allow the oven to cool until the fan stops.
2.
Remove the temperature probe.
3.
Fill in any hole left by the probe in the oven insulation.
4.
Replace the left-side instrument cover.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......