Clarus 600/680 User’s Guide
4-55
Setting up an FPD
WARNING
Flame Photometric Detectors (FPD) use hydrogen as fuel. If
the hydrogen is turned on without a column attached to the
detector fittings inside the oven, hydrogen could diffuse into the
oven creating the possibility of an explosion. To avoid possible
injury, DO NOT TURN ON THE HYDROGEN unless a column
is attached to the detector fittings inside the oven and all fittings
have been leak tested
.
Before disconnecting a column, make certain that the
HYDROGEN HAS BEEN TURNED OFF.
If two FPDs are installed and only one FPD has a column
connected to it, make certain that you plug the unused detector
inlet fitting with a
1/8-inch stainless steel plug (P/N N930-0061).
NOTE:
To prevent potential damage to the FPD in the manual pneumatics version, the
hydrogen needle valve has been factory-set to a very low setting. Set both the
hydrogen and air inlet pressures to 30 psig, then measure and set the flows as stated
in this chapter.
NOTE:
The PPC pneumatics detector gas modules have been set to OFF at the factory.
FPD Overview
In the Flame Photometric Detector (FPD), the column effluent is mixed with
hydrogen and then burned in air. Light emitted from the flame passes through a
lens, filter, and to a photomultiplier tube, which generates an electrical signal.
The Flame Photometric Detector is used for detecting sulfur, phosphorus, or tin
compounds, which produce chemiluminescent reactions with emissions at
wavelengths characteristic of the S
2
, Sn, and HPO species.
A filter that isolates sulfur emissions (P/N N6000637) is supplied with the detector.
Another filter that isolates phosphorus emissions (P/N N6000981) and a filter that
isolates tin emissions (P/N L4145472) are available from
PerkinElmer
at
www.perkinelmer.com
.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......