Using the Tools Menu
6-54
2.
In
PPC Reference Conditions
screen add the reference temperature and
pressure using the plus minus buttons. Touch
OK
when completed.
Calibrating the Carrier Gas Modules
CAUTION
If you are performing this calibration procedure and you do not
have a column connected to the detector, you must plug the
detector fitting in the oven with an Swagelok 1/8-inch cap before
proceeding. This will ensure that all the gas flows up through the
detector and not into the oven.
If you are using a packed injector or programmed on-column injector, you can
disconnect the column and measure the flow at the exit of the injector.
If you are using a capillary injector or programmed split/splitless injector, you
must remove the column. Cap the injector exit in the oven and measure flows at
the split vent exit.
To cap the injector exit in the oven, locate the Capillary Column Shipping Kit
(P/N N6100160) and insert the solid ferrule (P/N 09920511) into the 1/16-inch
nut (P/N 09903392) and connect the nut to the exit of the injector.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......