Using the Tools Menu
6-60
Calibrating the Oven Temperature
The Clarus GC oven temperature is controlled by a sophisticated software routine
that permits adjustment of the actual oven temperature inside the oven up or
down within a
±
10
°
C window. This calibration utility is intended for
applications which require precisely matching oven temperatures across several
Clarus GCs, such as the determination of retention indices. The calibration
information is stored in a reserved area of the battery-backed memory.
Required Equipment:
A calibrated precision platinum resistance
thermometer with 0.01
°
C readout and a probe capable of insertion into the GC
oven.
The following steps summarize how to calibrate the oven temperature:
1.
Calibrate the reference thermometer
2.
Place the thermometer probe in the oven
3.
Equilibrate the oven temperature
4.
Enter the required offset value
5.
Remove the thermometer probe
Calibrate the Reference Thermometer
Calibrate your reference thermometer according to its instructions. If the
thermometer has been calibrated by an outside service, there is no reason to
recalibrate it yourself.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......