Clarus 600/680 User’s Guide
4-31
Equalize the Makeup and Reference Gas Flows
The following procedures describe how to adjust the makeup and reference gas
using PPC pneumatics in the Clarus GC.
CAUTION
The TCD flow module for reference gas is calibrated with
helium at the factory. You must recalibrate the TCD flow
module if you are using a reference gas other than helium.
For optimum performance, apply a minimum flow of 5 mL/min to each of the two
detector channels.
Set the Makeup and Reference Gas
With PPC pneumatics the reference gas is set from the Clarus GC touch screen.
Also, PPC pneumatics provide you with TCD filament protection by monitoring the
carrier gas leak tests or monitoring the gas through the alarm feature and shutting
down the TCD current if a gas leak or no gas at all is detected. The PPC modules are
calibrated at the factory; therefore, to set the flows for an analysis, simply follow
this procedure.
1.
To set the make-up and reference gases touch the
Tools
button and touch
Configuration
from the drop down menu.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......