Clarus 600/680 User’s Guide
3-31
We recommend that if you are operating with a split vent flow of 50 mL/min or
less, select the
Auto
mode.
Before each injection in the
Auto
mode, the system briefly closes the split vent
and measures the total flow. With the split vent off this flow is the column flow
plus the septum purge flow. The septum purge flow is maintained due to a
mechanical restrictor and the split vent flow and split ratio are maintained for
each injection.
In the
Fixed
mode, you enter an offset value which is equal to the septum purge
flow plus the column flow at the time of injection. Since the system controls the
total flow (equal to the split vent plus septum purge plus column flow), the
septum purge is fixed, and the column flow is controlled, the difference between
these two flows and the offset entered will go out the vent flow. This will be
consistent for each run.
5.
If
Fixed
is selected use the up and down arrow or keypad
buttons
to enter an offset value in mL/min.
6.
Press
OK
.
Skip steps 9-16 if you are only programming the capillary control for pressure.
7.
If you wish to program the
Capillary Control
for more than just pressure, check
the
Capillary Control
check box from the
Injector
tab.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......