Clarus 600/680 User’s Guide
4-51
Activating the NPD Bead
The bead requires an activation step prior to use.
Do not inject a sample during
Activation.
To Activate the bead, carefully follow this procedure in sequential steps below:
1.
Turn the bead potentiometer dial clockwise to apply current to the bead.
2.
Slowly increase the NPD control voltage potentiometer to a setting of 700.
Note:
The bead will gradually acquire an orange glow at this point.
3.
Wait for 5 minutes then increase the potentiometer to 750.
4.
After 1 minute move the potentiometer to 800.
5.
After 1 minute move the potentiometer to 850.
6.
After 1 minute move the potentiometer to 900.
7.
After 1 minute reduce the potentiometer to 850.
8.
After 1 minute reduce the potentiometer to 800.
9.
After 1 minute reduce the potentiometer to 750.
10.
After 1 minute reduce the potentiometer down to 700
11.
For Capillary columns:
Wait 10 minutes, go to the NPD detector screen,
then adjust the potentiometer so that the NPD baseline response reads
around 0.5 mV on Range 1. Then set the desired column flow for the
application to be run.
For Packed Columns:
12.
Allow the baseline to stabilize for 30 minutes.
Wait 10 minutes, go to the NPD detector screen,
then adjust the potentiometer down to 500 or below to drop the bead below
the active level. Remove the Detector Cap Nut (P/N N9300061) and
reconnect the column to the detector. Set the desired column flow for the
application to be run, then again increase the bead current potentiometer to
give a reading around 0.5 mV on Range 1.
13.
Using the potentiometer, adjust the baseline between 0.4 and 0.6 mV at
Range 1 and the bead ready to use.
14.
After conditioning the bead remove the Swagelok plug from the detector
fitting and reconnect the column.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......