Clarus 600/680 User’s Guide
4-33
3.
In the TCD Configuration screen set the
Make-up gas
and
Reference gas
line
flows to
30
mL/min using the up and down arrow buttons or the numeric
keypad.
4.
As shown in the example above, use the up and down buttons or the numeric
keypad to set the filter to
200
and the Output to
5
.When all the settings are
appropriate for your analysis touch
OK
. The TCD screen will now show the
changes you made in the configuration settings for the TCD detector.
The default setting for the filter is 200 milliseconds. If you want to improve
the signal-to-noise, change the setting to 800 milliseconds; however, the noise
improves but the peak shapes broaden. If you want a faster detector response,
change the setting to 50 milliseconds; however, the noise level may increase.
NOTE:
To protect the TCD filaments from errors associated with the sample and reference
gas channels, you must select the necessary detector gas sources. As shipped from
the factory, the sample channel is Carrier 2 and the reference gas channel is the
detector Reference Gas. In some cases, with Arnel configurations, for example, the
TCD could be sourced with both a Carrier gas as well as an Aux gas (in GSV
applications half the run could look at the Carrier and half could look at an Aux
source). In other cases, the reference gas could be replaced by a second channel of
carrier gas (in a dual column TCD setup).
Enter the TCD Bridge Current Range
Select a bridge current range using the criteria in the following table.
NOTE:
Operating at too high a temperature can have an adverse effect on filament life.
Table 4-1. Bridge Current Range vs. Max. Detector Operating Temperature
Bridge Current Range
Recommended Operating Temperatures
for the Following Detectors
He or H
2
N
2
Argon
±
4
Up to 100
°
C
Do not Use 4
Do not Use 4
±
3
Up to 300
°
C
Do not Use 3
Do not Use 3
±
2
Up to 350
°
C
Up to 110
°
C
Do not Use 2
±
1
Up to 350
°
C
Up to 350
°
C
Up to 350
°
C
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......