Setting Up the Detectors
4-6
Setting Up an FID
To set up an FID that uses PPC modules, first read the FID Setup Summary, then
follow these procedures.
WARNING
To avoid injury, Do Not
Flame Ionization Detectors (FID) use hydrogen gas. Before
you turn on the hydrogen gas to the FID, ensure that a
column is connected to the FID fitting inside the oven.
Turning on the hydrogen gas without first connecting a
column to the FID could create the possibility of an
explosion in the oven.
turn on the hydrogen gas unless a
column is connected to the FID and all fitting connections
have been leak tested
.
WARNING
Turn off the hydrogen gas before disconnecting a column
from the FID.
When two FIDs are installed, but only one FID has a
column connected to it, plug the inlet of the FID without a
column with a 1/8-inch stainless steel plug (P/N N930-
0061) to prevent hydrogen from back-diffusing into the
oven
.
Since in the manual pneumatics version, the hydrogen
pressure regulator and the air needle valve have been
turned off prior to shipment, first set the hydrogen and air
inlet pressures to 30 psig. Then measure and set the gas
flows as described in this chapter.
NOTE:
The detector gas modules have been set to OFF at the factory for both manual and
PPC pneumatics.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......