Clarus 600/680 User’s Guide
4-53
NOTE: (
1) The bead will be stable after operating for several hours, but it will drift in time.
At the start of each day, adjust the background reading to the setting you are using
and allow a few minutes for it to stabilize.
(2)
Due to the loss of alkali metal with use, the nature of the bead is to drift over time.
We strongly recommend using an internal standard for quantitative analysis.
(3) The bead can operate at higher background settings. The higher the setting, the
greater the signal and noise. Therefore, the signal to noise ratio will not increase
dramatically. Operate at the lowest possible setting to achieve the required sensitivity;
this will also prolong the life of the bead.
NOTE:
The bead should be turned off by turning the bead potentiometer dial fully
counterclockwise before you turn off the Clarus GC. Remember that the bead has a
finite life. You can extend the bead life by turning the bead off when it is not in use
for long periods of time (for example, over the weekend).
Reactivate a Used NPD Bead
To increase the bead life, turn off the bead if the NPD is not continually in use.
When you turn off the bead, keep the detector temperature, hydrogen flow, and air
flow set to operating conditions. If the column remains connected to the injector
and detector in the oven, make sure the carrier gas is flowing through the column. If
the column is disconnected from the detector, make sure the Swagelok cap is
installed on the NPD fitting.
Prior to using the NPD, adjust the bead background to the same millivolt setting that
you were using. Run your standards to determine if the NPD bead is responding at a
level of sensitivity equal to the previous run. If not, then reactivate the NPD bead.
To reactivate a NPD bead follow the activation procedure above.
Operating Notes
The bead is operated as a conventional NPD bead; following these instructions will
increase the bead lifetime.
Baseline:
Set the baseline to a level of between 0.4 – 0.5. A lower baseline will
help to extend bead lifetime. Studies have shown that the signal-to-noise will not be
adversely affected by the lower baseline.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......