Using the Active Method
3-46
avoid this, the setup procedure is slightly different than the above procedure. When using
a mass spectrometer, you would not want the initial pressure to be too high, because at
the end of the GC run all parameters are set back to their initial values, and the higher
pressure could cause damage to the vacuum system.
For the Clarus GC/MS the initial value is set to the normal operating pressure (flow or
velocity) and a negative timed event with a high pressure (flow or velocity) will be added
to the method. This causes the pressure (flow or velocity) to increase prior to Ready (for
example, prior to the injection). At injection, the GC method holds this pressure (flow or
velocity) for the time necessary to transfer the sample onto the column. The
experimentally determined pressure (flow or velocity) is then programmed at a controlled
rate down to the initial (or normal) column operating pressure (flow or velocity).
If you are using the autosampler, at the end of the last sample in the sequence, the Clarus
MS will go “Not Ready,” thereby sending a signal to the GC keeping it from coming
Ready. (The External Ready must be connected. See the
Clarus 600/680 GC Installation
Instructions, 09936779
.) This keeps the GC from executing the negative timed event and
holds the pressure at the lower setting.
Setting the Split Flow for Splitless Injection
To make a splitless injection, initially turn Off the appropriate split flow that is associated
with your injector with a timed event. Then with another timed event, turn it On from
0.25 to 2 minutes into the GC run and leave it on. When the GC run is complete, the GC
returns the split flow to its initial setting.
To set up for splitless operation:
From the Injector status screen touch the dialog box for
Split
flow and use the up
and down arrow or keypad
buttons to enter a split flow in mL/min.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......