Using the Active Method
3-22
Initial Carrier-Gas Setpoint
The carrier gas mass-flow controller receives set points of flow or pressure depending on
the configured carrier gas mode. The program calculates the initial set point from the
active GC method. The initial set point is established upon setting up a method, ending a
run, or resetting the GC. The flow rate is held constant at the initial value if no additional
carrier-flow entries are made in the GC method. The controller receives pressure set
points when operating in the pressure control mode.
Carrier-Gas Program
The GC method accommodates up to three additional carrier-gas program steps each
consisting of a hold time, ramp rate, and plateau set point. Sequential set points may be
greater than or less than the preceding set point. The PPC system programs the carrier-
gas controller with a positive or negative ramp as required.
At the end of the GC oven program, the carrier gas resets to the initial set point value in
the method even if a carrier-gas program is in progress. The carrier gas also resets if the
end of the carrier-gas program occurs before the end of the oven program.
Carrier-Gas Timed Events
You can also use timed events to produce step changes in the set point. Carrier-gas events
can occur before or after a run starts. A carrier-gas event supersedes the method set point
value at the time the event occurs. However, if the carrier-gas program in the method
advances to a new step after a carrier-gas event, the new program step or ramp takes
control.
Leak Checking
The Clarus GC monitors the inlet pressure with a carrier gas mass-flow controller in use.
After selecting the carrier-gas
Flow
mode in configuration, you can enter the maximum
inlet pressure value. If the inlet pressure drops below or rises above the set limit for more
than one minute, a warning message appears.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......