Setting Up the Detectors
4-14
NOTE:
For proper ignition to occur, the FID must be heated to at least 100
°
C.
2.
Touch the temp field
and use the up and down arrow or keypad
buttons to enter a setpoint of at least 100
°
C, or enter a value 50
°
C
higher than the highest temperature of your oven program.
3.
Wait for the FID temperature to become READY. When the temperature is
ready a green light appears next to the temperature display
.
4.
Touch the
Ignite
button
.
(Note the signal level in mV before the flame ignites.)
5.
Upon pressing the Ignite button, the air flow turns off for 15 seconds,
Flame
Ignition 1
appears in the status bar, and the glow plug begins to glow.
Glow Plug Ignitor
Figure 4-2 Location of the FID glow plug ignitor.
6.
When you hear a "
POP
,” observe the
Signal
level in
mV
as it increases on the
screen and remains greater than the previously noted value.
After the flame is lit, a screen will appear that shows the signal level at the range
that you set. If you have set a range of 20, the signal will be 20 times smaller.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......