Clarus 600/680 GC User’s Guide
6-71
The carrier gas flow rate is in units of mL/min (1 to 100mL/min in 0.1mL/min
increments). There is no default as this field will be already filled by a previous
calculation result.
Velocity
This is a selectable field and, if selected, the t0 field is also enabled for input. The
Inlet Press and Flow fields are deselected and are display only.
The carrier gas velocity is in units of cm/s (1 to 500cm/s in 0.1 cm/s increments).
There is no default as this field will be already filled by a previous calculation
result.
Split Flow Calculations Screen
Column flow rate
This field will take its initial value from the Main Screen Flow field. The same
range and units will apply.
If the value is changed in this field, the entry in the Main Screen will also be
updated and the input pressure, velocity and t0 will be recalculated.
Split flow rate
This field will use mL/min units (0 to 500mL/min range in 1ml/min increments).
The default will be 50mL/min.
Column i.d.
This field will take its initial value from the Main Screen Diameter field. The
same range and units will apply.
If the value is changed in this field, the entry in the Main Screen will also be
updated and the active input pressure, flow or velocity and t0 will be
recalculated.
Liner i.d.
This is the internal diameter of the injector liner. The input range will be 0.1 to
6.0mm in 0.1mm increments.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......