Clarus 600/680 User’s Guide
4-29
TCD Setup
Setting up a TCD requires that you equalize carrier gas flows through both channels,
followed by balancing the current through the two filaments. The TCD is
nondestructive and can be used in series with other detectors. The TCD is a
universal detector that is concentration sensitive. The TCD sensitivity depends on
the filament current, component concentrations, and the difference in thermal
conductivity between the carrier gas and the component.
NOTE:
To prevent TCD filament damage, shut off the filament current before
disconnecting columns or turning off gas flows.
To set up a TCD, read the TCD setup summary, then follow the steps.
TCD Setup Summary
To set up a TCD:
•
Ensure that a column is installed.
•
Turn on the oven and detector heaters.
•
Equalize the makeup and reference gas flows.
•
Enter the TCD bridge current.
•
Balance the bridge
Turn on the Oven and Detector Heaters
1.
Close the oven door. This turns on the oven heater.
2.
From the System Status screen touch the TCD button
.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......