Clarus 600/680 User’s Guide
3-49
8.
From the
Event
drop down menu select
Split A
(if you have the injector
configured for position A) or
Split B
(if you have the injector configured for
position B).
9.
Touch the
Time
dialog box and use the up and down arrow or keypad
buttons to enter a positive time for the event (for example, 2.00).
NOTE
:
The "Time" refers to the time (in minutes) into the GC run that you want to automatically
turn the valve on. The time depends on your sample, solvent, column, etc. Typical time
values are between 0.25 and 2.0 minutes. An optimum "Time" is the time that provides
maximum area counts (indicated by your data handling device) with minimum solvent
effect (tailing).
10.
Touch the
Value
dialog box and use the up and down arrow or keypad
buttons to enter a value of 50 mL/min.
At the end of a run, the split flow resets to the original setting. Setting a Value of
50 mL/min minimizes contamination by keeping the splitter on during cool-
down.
NOTE:
If you wish to conserve carrier gas, set the initial flow to 0.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......