Clarus 600/680 User’s Guide
4-3
Overview
This section contains procedures for setting up the following detectors:
FID (flame ionization)
PID (photoionization)
ECD (electron capture)
NPD (nitrogen phosphorus)
TCD (thermal conductivity)
FPD (flame photometric)
The instructions provided in this chapter describe how to set detector gases that are
controlled by PPC modules.
CAUTION
The Clarus GC PPC modules are calibrated at the factory with
their corresponding gas; however, there are exceptions. The
PPC module that controls the:
•
ECD is calibrated with nitrogen.
•
TCD is calibrated with helium.
•
PID is calibrated with helium.
If you intend to use a gas that is different from the gas used to
calibrate the PPC module, you must recalibrate the PPC
module with the gas.
The instructional procedures in this chapter are written with the assumption that:
•
You know how to measure gas flows using a flowmeter or with the Clarus
GC’s built-in stopwatch and a soap-bubble flowmeter. If you are not
familiar, prior to continuing, please read "Using the Built-in Stopwatch" in
Chapter 4, "Before You Install a Column."
•
All heaters are turned
OFF
and cool to the touch.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......