Using the Active Method
3-8
NOTE:
When using a POC or PSS injector with the subambient accessory, the injector
temperature must not be set below the oven temperature. You can minimize coolant
consumption if using the oven subambient with the PSS or POC installed but not being
used, by deconfiguring the PSS or POC.
Setting a Coolant Cut-In Temperature
The coolant cut-in temperature is the temperature at which the coolant starts to cool the
oven. The default cut-in temperature is 60
°
C. The allowable cut-in temperature range is
30
°
C
-
110
°
C. We recommend that
YOU NOTE THE AMBIENT TEMPERARTURE
AND SET
the cut-in temperature 20 °C above your current ambient temperature.
For example: A room temperature of 20
°
C uses a default temperature of 40
°
C, whereas
a room temperature of 25
°
C uses a 45
°
C cut-in temperature. Examples for setting up
isothermal, 1, and 2 ramp programs follow.
NOTE:
Higher cut-in temperatures cause more coolant to be consumed.
Setting Up an Isothermal Program
Setting up an isothermal program requires entering the following information in the
sequence shown:
•
Initial oven temperature (Temp 1)
•
The total run time for the initial temperature (Time 1)
•
A temperature rate increase of 0 (Rate 1).
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......