Using the Active Method
3-62
This option instructs the autosampler to interrupt the vial injection sequence and inject a
"Priority" vial that has been placed in the "Priority" position of the turntable. The priority
vial will be injected after completion of the current run.
After the priority vial has been injected (in accordance with the current program
parameters), and the priority run completed, the program continues with the next
injection in sequence.
NOTE:
Priority can only be selected when with the above screen is visible.
About Waste and Wash Vials
The Autosampler software has been optimized to use waste vials and solvent wash vials
to minimize contamination of the syringe needle.
Septa are available for the waste vials and solvent wash vials by ordering P/N N930-
2780. We recommend using these septa in place of the diffuser caps when you are using:
•
Very viscous samples (to wipe the sample off the outside of the syringe needle).
•
Toxic solvents and samples.
•
Very volatile solvents.
Multi-Program Mode
Waste Vials
When operating in the Multi-Program Mode, the sample pre-washes are emptied into
waste vial #1 for program 1 and waste vial #3 for program 2. The solvent post-washes
are emptied into waste vial #2 for program 1 and #4 for program 2.
Wash vials
Program 1 splits between solvent wash vials #1 and #2. For example, if four washes are
required, two will be taken from vial #1, and two will be taken from vial #2.
Program 2 splits between solvent wash vials #3 and #4. For example, if five washes are
required, three will be taken from vial #3, and two will be taken from vial #4.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......