Using the Active Method
3-54
2.
In the next screen touch the
Autosampler
radio button to the right of the
Type
heading.
The Autosampler screen contains three tabs. The first two tabs display identical screens
for setting
Program 1
and
2
, allowing you to sequence two methods. The
Tasks
tab
displays buttons to perform tasks such as Park, Clean, and specify a priority injection.
All of the screens provide status information.
The method field allows the user to run the GC under the control of a method other
than the current active method.
Start Button
Touch the Start button to start a run. The Start button is active whether or not the system
is ready. Issuing the start command before the system is ready gives users the flexibility
to set up and run a method without having to wait for the system to attain ready status. As
soon as you touch
OK
, the Start button changes to
Stop
whether or not the system is
ready. This enables you to cancel the run before it starts.
•
If the system is under external control (e.g. from TotalChrom) and a method has not
yet been set up, the Start Button is inactive.
•
In all other instances, even though the GC is not Ready, the Start Button will start a
run.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......