Setting Up the Detectors
4-68
4.
On this page touch the FPD button
to display the FPD Configuration
page.
5. In the PMT Voltage field set the percentage (see note below) by using the up
and down arrow or keypad
buttons. This will turn on the
photomultiplier tube.
NOTE:
The voltage to the photomultiplier tube has been set in the factory to approximately
70%. You can change this setting from the Clarus GC touch screen. You may want
to change this setting as the photomultiplier tube ages and the signal diminishes.
The signal-to-noise ratio may vary with each photomultiplier tube; therefore,
increasing the voltage may not increase the signal-to-noise ratio.
After installing a replacement photomultiplier tube, the noise should be about 2%
when the Clarus GC integrator output is set to attenuation 8 (x8) and the
Linearizer
off. See the
Clarus 600/680 GC Customer Hardware and Service Guide
"Maintenance” chapter, for this procedure.
NOTE:
The photomultiplier tube can be damaged if it is exposed to light while voltage is
ON. Never remove the detector cap or Photomultiplier tube WITH the voltage on.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......