Clarus 600/680 GC User’s Guide
6-49
Zeroing and Calibrating the PPC Modules
The Clarus GC gas modules have been calibrated at the factory and are ready to
use. For good laboratory practices and your own internal validation protocol,
you may wish to recalibrate at installation with your own flow measuring device.
All carrier gas modules have been calibrated at the factory with helium. If you
are using a different gas, you must enter the appropriate gas and recalibrate.
Detector gas modules must be recalibrated if you are using a different gas in your
laboratory. In the factory, the TCD reference gas and the PID make-up gas have
been calibrated with helium and the ECD make-up gas with nitrogen.
Zeroing and calibrating PPC modules consists of the following steps:
•
Zeroing the PPC gas modules (Gas turned off)
•
Setting the reference temperature and pressure
•
Calibrating the PPC carrier gas modules and/or PPC detector gas
modules (Gas turned on)
The calibration information is stored in a reserved area of the battery-backed
memory.
Zeroing the PPC Gas Modules
Before zeroing the PPC gas modules:
•
Make sure that the Clarus GC has been properly installed and all gases are
connected.
•
Turn on the Clarus GC power and
allow the GC to warm up for
approximately 15 minutes
before calibrating any gas.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......