Using the Tools Menu
6-70
Name
The name of the carrier gas is selected in order to apply the correct viscosity
coefficients and suggested column carrier gas velocities as shown in the Table.
Gas
Screen Name
ref
η
,Pa.s
x
Suggested
Velocity, cm/s
Helium
He
0.000018662
0.646
20 to 40
Hydrogen
H2
0.000008399
0.725
30 to 55
Nitrogen
N2
0.000016736
0.680
8 to 16
The default will be “He”.
Outlet Press
The column outlet pressure is set to Ambient or Vacuum (0) or to a user entered
Other value. Units will be in absolute units (psi abs or kPa abs) and will be set
according to the units chosen in the Configuration screens for the rest of the user
interaction. The range for this field will be 0 to 100 psi abs in 0.1psi increments
or 0 to 689 kPa abs in 1kPa increments. The default will be Ambient (14.7psi abs
or 101kPa abs).
Suggested Velocity
A display-only field that shows the recommended carrier gas velocity range from
the selected carrier gas name.
Inlet Press
This is a selectable field (this is the default selection) and, if selected, the Flow
and Velocity/t0 fields are deselected and are display only.
The carrier gas input pressure is in units of psig (1 to 100psig in 0.1psi
increments) or kPa (1to 689kPa in 1kPa increments). The default is 5.0psig or
34kPa.
Flow Rate
This is a selectable field and, if selected, the Inlet Press and Velocity/t0 fields are
deselected and are display only.
Summary of Contents for CLARUS 600 GC
Page 1: ...CLARUS 600 680 GC User s Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus 600 680 GC User s Guide ...
Page 10: ...viii ...
Page 11: ...Introduction 1 ...
Page 12: ...Introduction 1 2 ...
Page 23: ...Touch Screen Navigation 2 ...
Page 24: ...Touch Screen Navigation 2 2 ...
Page 73: ...Using the Active Method 3 ...
Page 74: ...Using the Active Method 3 2 ...
Page 87: ...Clarus 600 680 User s Guide 3 15 The injector information displays ...
Page 119: ...Clarus 600 680 User s Guide 3 47 Setting the Timed Events 1 Touch the Events tab ...
Page 137: ...Setting Up the Detectors 4 ...
Page 138: ...Setting Up the Detectors 4 2 ...
Page 194: ...Setting Up the Detectors 4 58 2 The FPD page is now visible ...
Page 199: ...Clarus 600 680 User s Guide 4 63 4 The FPD tab page is now displayed ...
Page 230: ...Setting Up the Detectors 4 94 NOTE When Autozero is Off the output signal is never autozeroed ...
Page 231: ...Using the Method Editor 5 ...
Page 232: ...Using the Method Editor 5 2 ...
Page 248: ...Using the Method Editor 5 18 ...
Page 249: ...Using the Tools Menu 6 ...
Page 250: ...Using the Tools Menu 6 2 ...
Page 272: ...Using the Tools Menu 6 24 PID NPD ...
Page 273: ...Clarus 600 680 GC User s Guide 6 25 FPD Output Configured ...
Page 282: ...Using the Tools Menu 6 34 PPC Configure ...
Page 329: ...Setting up a Typical Analysis 7 ...
Page 330: ...Setting up a Typical Analysis 7 2 ...
Page 333: ...Clarus 600 680 GC User s Guide 7 5 ...
Page 354: ...Setting up a Typical Analysis 7 26 ...
Page 355: ...Index ...
Page 356: ...I 2 ...
Page 361: ......